World class 3000, Instruction manual – Emerson Process Management 3000 User Manual

Page 175

Advertising
background image

Instruction Manual

Appendix J Rev. 1.1

June 1995

Rosemount Analytical Inc. A Division of Emerson Process Management

Appendices J-11

World Class 3000

DOES

HART UNIT

HAVE

POWER

?

DOES

HART UNIT

INDICATE

BATTERY

POWER

GOOD

?

IS

HART UNIT

BEING USED

"ON-LINE"

?

IS

HART UNIT

CONFIGURED

FOR IFT?

IS IFT

POWERED

?

IS IFT

EXPERIENCING

FAULTS

?

DOES

IFT HAVE

MICROPROCESSOR

BOARD

REV.#

3039513G

?

CLEAR IFT OF ALL FAULT

CONDITIONS.

YES

YES

YES

YES

YES

YES

YES

(GO TO

SHEET 2 OF 2)

REPLACE BATTERIES. IF

APPLICABLE, RECHARGE

NiCad BATTERIES.

VERIFY HART IS

CONFIGURED FOR IFT.

REFERENCE OFF-LINE

CONFIGURATION (HART

POCKET-SIZED MANUAL).

RECONFIGURE HART UNIT

FOR IFT.

CONTACT SERVICE

REPRESENTATIVE.

SUPPLY POWER TO IFT.

NO

NO

NO

NO

NO

NO

NO

686004

Figure J-5. Model 275D9E, Troubleshooting Flowchart (Sheet 1 of 2)

J

Advertising