SERVICE AND NORMAL MAINTENANCE

E-5 OVERVIEW

This section describes service and routine
maintenance of the Intelligent Field Transmitter.
Replacement parts referred to are available
from Rosemount. Refer to Replacement Parts
later in this appendix for part numbers and or-
dering information.

Install all protective equipment covers
and safety ground leads after equip-
ment repair or service. Failure to in-
stall covers and ground leads could
result in serious injury or death.

E-6 FUSE

REPLACEMENT

Power supply board (4, Figure E-6) contains
four identical 5 amp fuses.

Perform the following procedure to check or re-
place a fuse. In addition, 2 additional 5 amp
fuses (F1 and F2) are included if the IFT unit
has an internal heater installed.

Disconnect and lock out power before
working on any electrical components.
There is voltage up to 240 Vac, and
could cause personal injury.

a. Turn off power to the system.

b. Open cover door (16) of the IFT by remov-

ing screws (17).

c. Remove protective cover (13) by removing

screws (14) and washers (15).

d. Unscrew fuseholder top and remove the

fuse (5). After checking or replacing a fuse,
reinstall the fuseholder top.

e. Replace protective cover (13) and secure

with washers (15) and screws (14).

f. Close cover door (16) and secure with

screws (17).

E-7 TRANSFORMER REPLACEMENT

Disconnect and lock out power before
working on any electrical components.
There is voltage up to 240 Vac, and
could cause personal injury.

a. Turn off power to the system.

b. Open cover door (16) of the IFT by remov-

ing screws (17).

c. Remove protective cover (13) by removing

screws (14) and washers (15).

d. Disconnect cable (1) from the receptacle on

microprocessor board (11). Disconnect GUI
assembly cable from receptacles on micro-
processor board if IFT is equipped with GUI.

e. Carefully tagging wires, remove the wires

from terminal strip on interconnect board
(12).

f. If unit is equipped with heater option, re-

move thermoswitch assembly (18, 19,
Figure E-8) by removing screws (13) and
washers (14).

g. Remove mounting plate (10) by removing

the necessary screws.

h. Disconnect transformer cable plugs from the

receptacles on power supply board (4).

i. Remove transformer (9) from enclosure (6)

by removing four screws (8).

j. Attach new transformer to enclosure (6) with

four screws (8).

k. Connect the transformer cable plugs from

transformer (9) to the receptacles on power
supply board (4).