World class 3000, Installation, Instruction manual – Emerson Process Management 3000 User Manual

Page 167

Advertising
background image

Instruction Manual

Appendix J Rev. 1.1

June 1995

Rosemount Analytical Inc. A Division of Emerson Process Management

Appendices J-3

World Class 3000

INSTALLATION

J-4

HART COMMUNICATOR SIGNAL LINE
CONNECTIONS

The HART Communicator can connect to the
IFT analog output signal line at any wiring ter-
mination point in the 4-20 mA current loop.
There are two methods of connecting the HART
Communicator to the signal line. For applica-
tions in which the signal line has a load resis-
tance of 250 ohms or more, refer to method 1.

For applications in which the signal line load re-
sistance is less than 250 ohms, refer to
method 2.

a. Method 1, For Load Resistance > 250

Ohms

Refer to Figure J-2 and the following steps
to connect the HART Communicator to a
signal line with 250 ohms or more of load
resistance.

R

250

L

>_

W

LOOP CONNECTORS

USE INTERFACE
00275 0013 ONLY

SERIAL PORT & BATTERY
CHARGER MUST
NOT BE USED IN
HAZARDOUS AREAS

SERIAL PORT

INTELLIGENT FIELD

TRANSMITTER IFT 3000

J1

JM1

JM6

J2

J3

J4

J5

J6

J7

J8

J9

I

V

MICROPROCESSOR

BOARD

CURRENT/VOLTAGE SELECTION SWITCH

TO "CURRENT" POSITION

INTERCONNECT BOARD

ANALOG OUTPUT

4-20mA SIGNAL LINE

ANALOG OUTPUT DEVICE

LEAD SET

HART

COMMUNICATOR

HART COMMUNICATOR

REAR PANEL

LOOP CONNECTORS

686002

Figure J-2. Signal Line Connections, > 250 Ohms Load Resistance

J

Advertising