Signal Connections

Signal connections are made to a nine-terminal connector labeled
SIGNAL on Figure 5. The COM (common) terminal is the return for the
4-20 mA and the

±10 V input signals.

Current and Voltage Connections

The following figures show examples of current and voltage connections.
There are no switches or jumpers to set up for current and voltage inputs.
Setup and programming is performed through the front panel buttons.

Figure 7. Transmitters Powered by Internal Supply

Figure 8. Transmitter Powered by Ext. Supply or Self-Powered

The current input is protected against current overload by a resettable
fuse. The display may or may not show a fault condition depending on
the nature of the overload.
The fuse limits the current to a safe level when it detects a fault
condition, and automatically resets itself when the fault condition is
removed.