SECTION 2 - INSTALLATION

2.1

UNPACKING AND INSPECTION

The Model TMA 4882-27 has been carefully inspected and tested prior to packing. Inspect the
shipping carton upon receipt for evidence of damage during transit. Save the original packing
material. If any indication of damage is found, file a claim immediately with the responsible
transport service.

For repairs of a product damaged in shipment, contact the Kepco Factory Representative near-
est you or the Kepco Sales Department directly for further instruction.

2.2

INSTALLATION

The installation and set-up procedure for the TMA 4882-27 consists of the following steps:

1. Set Device Address Selector and/or configure RS 232 port (PAR. 2.2.1).

2. Select start-up language (SCPI or CIIL), GPIB addressing (Primary/Secondary) and Com-

patibility Mode (PAR. 2.2.2).

3. Set Shield Ground Jumper (PAR. 2.2.3).

4. Select GTL (Go To Local) command inhibit/enable (PAR. 2.2.4).

5. Perform final system interconnections (PAR. 2.2.5).

2.2.1

SET (GPIB) DEVICE ADDRESS, CONFIGURE RS 232 (SEE FIGURE 2-1)

A single set of DIP switches, accessible at the rear panel (Figure 2-1), are used both to set the
GPIB Device Address (the factory default is 6) and to configure the RS 232 port. The following
paragraphs explain how to proceed if using GPIB only (PAR. 2.2.1.1), RS 232 only (PAR.
2.2.1.2) or both (PAR. 2.2.1.3). RS 232 connections are explained in PAR. 2.2.1.4) and RS 232
operation is described in PAR. 3.5

The Device Address for the TMA 4882-27 is initially set by means of DIP switches accessible at
the rear panel (Figure 2-1). The Device Address is the permanent Listener or Talker address of
the TMA 4882-27 on the GPIB. It is factory preset to address 6. If a different Device Address is
required in your system, proceed as follows. There are 31 (0-30) possible choices (See Table 2-
1)
.

2.2.1.1

USING GPIB ONLY

The Device Address is the permanent Listener and Talker address of the MST 488-27 on the
GPIB. It is factory preset to address 6. If a different Device Address is required in your system,
proceed as follows. There are 31 (0-30) possible choices (See Table 2-1).

1. Place TMA 4882-27 power module controller with the rear panel facing you.

2. The Device Address DIP switches are positions A1 through A5 (from left to right). These

switches are preset by Kepco to address 6. For other device addresses set them according
to Table 2-1.