A.9 *opt? — options query, Figure a-1. gpib commands, A.10 *rst — reset command – KEPCO TMA 4882-27 User Manual

Page 49: A.10, Re a-1.), Gure a-1.), Ure a-1.), A-1.)

Advertising
background image

TMA4882-27 021910

A-3

A.9

*OPT? — OPTIONS QUERY

*OPT?

Syntax:

*OPT?
Returns string determined by power supply model.

Description: Causes the power supply to return an ASCII string which defines the functionality of the power

supply. The functionality is defined as follows:

FIGURE A-1. GPIB COMMANDS

A.10 *RST — RESET COMMAND

*RST

Syntax:

*RST

Description: Resets power supply to the power on default state. The power supply is programmed to the power

on values of the following parameters: CURR[:LEV][:IMM] = 0, VOLT[:LEV][:IMM] = 0, OUTP[:STAT]
= OFF. If the power supply is in either an overvoltage or overcurrent state, this condition is reset by
*RST. (See example, Figure A-1.)

STRING DATA

MEANING

RI1

Indicates unit has the ability to modify the DCL response.

CHN

Indicates multiple channel device is enabled

DS0

Display present, no text capability.

CIIL

Unit supports CIIL commands

*CLS

Power supply clears status data.

*ESE 60

Power supply enables bits 5, 4, 3 and 2, allowing command error, execution
error, device dependent error and query error to set the Event Status
Summary bit when an STB command is executed.

*ESE?

Returns 60, (value of the mask) verifying that bits 5, 4, 3 and 2 are enabled.

*ES

Unknown command will set command error (Bit 5).

*ESR?

Returns 32 (bit 5 set), indicating Command Error has occurred since the last
time the register was read.

*IDN?

Controller returns character string: “KEPCO,TMC,1,V3.0-3.0”

*OPC

Allows status bit 0 to be set when pending operations complete

VOLT 21;CURR 3

Sets output voltage to 21V, output current to 3A

*ESR

Returns 129 (128 + 1, power on, bit 7 = 1, operation complete, bit 1 = 1)

*ESR

Returns 0 (event status register cleared by prior *ESR?)

VOLT 15;CURR 5;*OPC? Sets output voltage to 15V, output current to 5A, puts “1” on output bus when

command operations are complete.

*RST

Power supply reset to power on default state.

*SRE 40

When ESB or QUES bits are set (Table A-3), the Request for Service bit will
be set.

*SRE?

Returns the value of the mask (40).

*STB?

For example, the Power supply responds with 96 (64 + 32) if MSS and the
Event Status Byte (Table A-3) summary bit have been set. The power supply
returns 00 if no bits have been set.

VOLT 25

Power supply voltage commanded to 25V.

VOLT:TRIG 12

Programs power supply voltage to 12V when *TRG received.

INIT

Trigger event is initialized.

*TRG

Power supply reverts to commanded output voltage of 12V.

*TST?

Power supply executes self test and responds with 0 if test completed
successfully, with 1 if test failed.