Conditions of Conformance

Programming Module

When this product is used in applications governed by the requirements of the EEC, the following restric-
tions and conditions apply:

1. For European applications, requiring compliance to the Low Voltage Directive, 73/23/EEC, this power

supply is considered a component product, designed for “built in” applications. Because it is incom-
plete in construction, the end product enclosure must provide for compliance to any remaining electri-
cal safety requirements and act as a fire enclosure. (EN61010-1 Cl. 6, Cl. 7, Cl.8, Cl. 9 and EN61010-
1 annex F)

2. This power supply is designed for stationary installation either within an equipment rack or a KEPCO

Rack Adapter RA 55 or CA 400.

3. This power supply is considered a Class 1 (earthed) product, and as such depends upon proper con-

nection to protective earth for safety from electric shock. (EN61010-1 Cl. 6.5.4)

4. This power supply is intended for use as part of equipment meant for test, measurement and labora-

tory use, and is designed to operate from single phase, three wire power systems. This equipment
must be installed in a specifically designed KEPCO rack adapter and within a suitably wired equipment
rack, utilizing a three wire (grounded) mains connection. See wiring section of this manual for com-
plete electrical wiring instructions. (EN61010-1 Cl. 6.5.4 and Cl.6.10.1)

5. This power supply has secondary output circuits that are considered SELV.

6. This power supply employs a supplementary circuit protector in the form of a fuse mounted within its

enclosure. The fuse protects the power supply itself from damage in the event of a fault condition. For
complete circuit protection of the end product, as well as the building wiring, it is required that a pri-
mary circuit protection device be fitted to the branch circuit wiring. (EN61010-1 Cl. 9.6.2)

7. Hazardous voltages are present within this power supply during normal operation. All operator adjust-

ments to the product are made via externally accessible switches, controls and signal lines as speci-
fied within the product operating instructions. There are no user or operator serviceable parts within
the product enclosure. Refer all servicing to qualified and trained Kepco service technicians.