Figure c-5. opn, cls — open, close relay commands, Figure c-6. rst — reset command, Opn, cls – KEPCO TMA 4882-27 User Manual

Page 70: Cnf, ist

Advertising
background image

C-4

TMA 4882-27SVC 021910

FIGURE C-5. OPN, CLS — OPEN, CLOSE RELAY COMMANDS

FIGURE C-6. RST — RESET COMMAND

FIGURE C-7. CNF, IST — CONFIDENCE TEST, INTERNAL SELF TEST COMMANDS

Syntax:

OPN :CHnn
CLS :CHnn

Function:

These commands are used to connect or disconnect the power supply from the load (effective for MR
and MGR options only).

Description:

OPN

Disconnects the load from the power supply specified by the operand.

CLS

Connects the load to the power supply specified by the operand.

Example:

OPN :CH22

Opens the relay of the power supply at node address 22.

CLS :CH14

Closes the relay of the power supply at node address 14.

OPN, CLS

Syntax:

RST DCS :CHnn

Function:

This operator is used to return a power supply to its power-on state. The output voltage and current
are programmed to zero and the output relay of MR and MGR models is opened.

Example:

RST DCS :CH13

The power supply at node address 13 is reset.

RST

Syntax:

CNF or IST

Function:

Causes power supply to execute confidence test.

Description:

The CNF operator commands the MBT to execute the confidence test procedure defined for the MBT
power supplies (IST is functionally identical to CNF for MBT power supplies. The procedure consists
of opening all power relays, programming voltage and current to their maximum values, switching
polarity, checking for error flags, then programming voltage and current to zero. The results of CNF
are obtained through the STA command.

Example:

CNF

All power supplies in the daisy chain execute confidence test.

IST

All power supplies in the daisy chain execute self test.

CNF, IST