APPENDIX C - CIIL COMMAND DEFINITIONS

C.1

INTRODUCTION

This appendix defines the CIIL commands used with the TMA 4882-27 Controller. Table C-1
provides a quick reference of all CIIL commands used in the TMA 4882-27.

FIGURE C-1. FNC — FUNCTION COMMAND

TABLE C-1. CIIL SUBSYSTEM COMMAND/QUERY INDEX

COMMAND

PAGE

COMMAND

PAGE

CLS

C-4

OPN

C-4

CNF

C-4

RST

C-4

FNC

C-1

SET

C-3

FTH

C-2

SRN

C-3

GAL

C-6

SRX

C-3

INX

C-2

STA

C-5

IST

C-4

Syntax:

Stimulus mode: FNC DCS:CHnn <SET Command>
Sensor mode:

FNC DCS <VOLT or CURR command> :CHnn

Function:

This operator is used with either the SET command to program a power supply's output (stimulus
mode), or with the VOLT and CURR commands to read its output settings (sensor mode).

Description:

The first operand contains the three (3) letter mnemonic pertaining to the device on the control bus, in
this case DCS (Direct Current Source). If a reading is being set up, the modifier VOLT or CURR fol-
lows. The next operand is used to select the specific channel of the device being programmed or read
from. The TMA 4882-27 can control up to 27 power supplies/power modules with control bus
addresses in the range of 1 to 31.

Example:

FNC DCS :CH12 SET VOLT 15

Power supply at node address 12 commanded to 15V

FNC DCS :CH12 SET CURR 3

Power supply at node address 12 commanded to 3A

FNC DCS VOLT :CH03

Power supply at node address 3 returns value which represents
actual output voltage

FNC DCS CURR :CH21

Power supply at node address 21 returns value which represents
actual output current

NOTE: Actual output voltage and current depends on whether output is enabled or disabled

and load conditions