World class 3000, Instruction bulletin – Emerson ROSEMOUNT 3000 User Manual

Page 118

Advertising
background image

Instruction Bulletin

Appendix A Rev. 3.9
May 2005

A-18 Appendices

Rosemount Analytical Inc. A Division of Emerson Process Management

World Class 3000

OPTIONAL CERAMIC

DIFFUSION ELEMENT

VEE

DEFLECTOR

HUB

PIN

WRENCH

SETSCREW

RETAINER

CEMENT

PORT

CEMENT

FILLET

21240026

Figure A-19. Removal of Optional Diffusor

and Vee Deflector

PROBE TUBE

(NOT INCLUDED

IN KIT)

SOCKET HEAD

CAP SCREWS

CORRUGATED

SEAL

CELL AND

FLANGE

ASSEMBLY

CALIBRATION GAS

PASSAGE

21240009

Figure A-20. Cell Replacement Kit

e. If contact assembly is damaged, replace

contact and thermocouple according to
paragraph A-13, Replacement of Contact
and Thermocouple Assembly.

f. Remove and discard corrugated seal. Clean

mating faces of probe tube and retainer.
Remove burrs and raised surfaces with
block of wood and crocus cloth. Clean
threads on retainer and hub.

g. Rub a small amount of anti-seize on both

sides of new corrugated seal.

h. Assemble cell and flange assembly, corru-

gated seal, and probe tube. Make sure the
calibration tube lines up with the calibration
gas passage in each component. Apply a
small amount of anti-seize compound to
screw threads and use screws to secure
assembly. Torque to 55 in-lbs (4 N

m).

i.

Apply anti-seize compound to threads of
cell assembly, hub, and setscrews. Reinstall
hub on cell assembly. Using pin spanner
wrenches, torque to 10 ft-lbs (14 N

m). If

applicable, reinstall vee deflector, orienting
apex toward gas flow. Secure with
setscrews and anti-seize compound.
Torque to 25 in-lbs (2.8 N

m).

j.

On systems equipped with an abrasive
shield, install dust seal gaskets, with joints
180

o

apart.

k. Reinstall probe and gasket on stack flange.

If there is an abrasive shield in the stack,
make sure dust seal gaskets are in place as
they enter 15

o

reducing cone.

l.

Turn power on to electronics and monitor
thermocouple output. It should stabilize at
29.3 ±0.2 mV. Set reference air flow at 2 scfh
(56.6 L/hr). After probe stabilizes, calibrate
probe per Instruction Bulletin applicable to
your electronics package. If new components
have been installed, repeat calibration after
24 hours of operation.

A-12 OPTIONAL CERAMIC DIFFUSION

ELEMENT REPLACEMENT

a. General

The diffusion element protects the cell from
particles in process gases. It does not nor-
mally need to be replaced because the vee
deflector protects it from particulate erosion.
In severe environments the filter may be
broken or subject to excessive erosion. Ex-
amine the diffusion element whenever re-
moving the probe for any purpose. Replace
if damaged.

Advertising