Instruction Manual

IB-106-300NH Rev. 4.3
May 2005

2-20 Installation

Rosemount Analytical Inc. A Division of Emerson Process Management

World Class 3000

ALWAYS DISCONNECT LINE VOLTAGE

FROM HEATER POWER SUPPLY AND

ANALOG ELECTRONICS (IF USED)

BEFORE CHANGING JUMPERS.

JUMPER

CONFIGURATIONS

LINE VOLTAGE

SELECTION

JUMPER

(INSTALL)

JUMPER

(INSTALL)

HEATER

POWER

JUMPER

100/120 VAC
220/240 VAC

JM4, JM1

JM5

REMOTE

ON

REMOVE JM2

*INSTALL JM2

ELECTRONICS

SELECTION

JUMPER

PROBE HEATER

VOLTAGE SELECTION

INSTALL JM3, JM6

REMOVE JM3, JM6

*ANALOG (EXISTING)

DIGITAL

(NEXT GENERATION)

WORLD CLASS PROBE

(44V)

JM7

JM8

218 PROBE (115V)

1

2

NOTES:

100 VAC OPERATION REQUIRES TRANSFORMER PART
NUMBER 1M02961G02.

REFER TO TABLE 3-5 FOR PROPER SET POINT SELECTION.

1

2

37840009

Figure 2-15. Jumper Selection Label

NOTE

Before supplying power to the heater
power supply, verify that jumpers JM3,
JM6 are removed and JM7 is installed.
If relay wire (Figure 2-13, note 1) is in-
stalled, JM2 must be removed from
HPS Mother Board (Figure 2-16).

4. Before supplying power to the heater

power supply, verify that the jumpers
on the mother board, Figure 2-16, are
properly configured. Jumpers JM3 and
JM6 should be removed and JM7
should be installed.

Additionally, make sure that the proper
jumper for your line voltage is installed,
Figure 2-15. If relay wire (Figure 2-13,
note 1) is not installed, JM 2 should be
installed on the HPS Mother Board
(Figure 2-16).

NOTE

Refer to Figure 2-8 and Figure 2-10 for
proper IFT jumper configuration. IFT
microprocessor and interconnect
board jumper configurations must be
set correctly in order for HPS to work
properly.

29850001

NOTE: ALL FUSES ARE 5A @ 250 VAC,

ANTI-SURGE, CASE SIZE
5 X 20 MM, TYPE T TO IEC127,
SCHURTER.

3D3

080G

REV

JM1

JM2
JM4

JM5

JM7

FUSE

FUSES

Figure 2-16. Jumpers on HPS Mother Board