Installation 2-9

World Class 3000

2-2

INTELLIGENT FIELD TRANSMITTER (IFT)
INSTALLATION

a. Mechanical

Installation

The outline drawing of the IFT module in
Figure 2-4 shows mounting centers and
clearances. The NEMA 4X enclosure is de-
signed to be mounted on a wall or bulk-
head. The IFT should be installed no more
than 1200 feet (364 m) from the optional
HPS or 150 feet (45 m) from the probe if
HPS is not installed in the system.

b. Electrical

Connections

To meet the Safety Requirements of
IEC 1010 (E
C requirement), and ensure
safe operation of this equipment, con-
nection to the main electrical power
supply must be made through a circuit
breaker (min 10A) which will discon-
nect all current carrying conductors
during a fault situation. This circuit
breaker should also include a me-
chanically operated isolating switch. If
not, then another external means of
disconnecting the supply from the
equipment should be located close by.
Circuit breakers or switches must
comply with a recognized standard
such as IEC 947.

NOTE

Refer to Figure 2-6 for fuse locations
and specifications.

1. The IFT can be configured for 100,

120, 220, or 240 line voltages. For 120
Vac usage, install JM8, JM7, and JM1
on the power supply board. For 220
Vac usage, install jumpers JM6, JM5,
JM2 (refer to Figure 2-5 and Figure
2-6).

2. For installations where the cable run is

less than 150 feet (45 m), the IFT can

LINE

PROBE

NOTICE:

ALL CABLE SHOULD
BE RATED FOR USE

ABOVE 76 C

Model IFT

Assy 6A00178GXX

SN XXXXXXXXXXXX

Volts XXX @ 50/60 Hz

Line Fuse 5 Amps
Rosemount Analytical Inc.

P.O. Box 901, Orrville, OH

44667 USA

8.00 (203.2)

13.10

(332.7)

6.40

(162.6)

11.0 (279.4) MINIMUM DOOR

SWING CLEARANCE

NOTE: DESIGN DIMENSIONS ARE IN INCHES

WITH MILLIMETERS IN PARENTHESES.

37840020

1.62

(41.1)

0.867

(22.00)

0

1.49

(37.8)

2.25

(57.15)

1.00

(25.4)

11.00 (279.4)

12.50

(317.5)

14.00

(355.6)

Figure 2-4. Outline of Intelligent Field Transmitter

be configured to connect directly to a
probe. An optional HPS is available for
cable runs over 150 feet (45 m). The
electrical connections for a non-HPS
equipped system should be made as de-
scribed in the electrical installation dia-
gram, Figure 2-7. Refer to Figure 2-13
for connections for an HPS equipped
system.

Advertising