5-14 General User Interface (GUI) Operation

Rosemount Analytical Inc. A Division of Emerson Process Management

World Class 3000

Table 5-5. SETUP Sub-Menu (continued)

SUB-MENU

SELECTION

PARAMETERS

DESCRIPTION

Analog Output

RANGE SETUP

(continued)

(Source set to Dual Rng O

2

)

Xfer Fnct

Log
Lin

Select the transfer function used on the analog output.
Selecting Log will not effect the output when Efficiency
is selected as the Source.

Normal Range Values

High End

0.000% O

2

- 25.00% O

2

Low End

0.000% O

2

- 25.00% O

2

Enter the upper and lower analog output range values
for Normal Operating Range. The High End value de-
fines the measured O

2

value corresponding to the high

analog output value, i.e, 20mA or 10V, and the Low
End value corresponds to the low analog output value,
i.e., 0mA, 4mA, or 0V.

Dual Range Setup

Mode Setup

Range Mode

Normal

Forces the output to the Normal Range.

Auto

Allows the IFT to select either the High Range or the
Normal Range based on the present O

2

value and the

Mode Setup Values.

High

Forces the output to the High Range.

High in Cal

Yes/No

Selecting Yes will cause the High Range to be used
whenever the probe is being calibrated.

Switches at
0.000% O

2

- 25.00% O

2

Enters the switching point between the High and Nor-
mal Ranges. O

2

values above this point will use the

High Range and values below this point will use the
Normal Range. The O

2

value must be below the switch

point by 10% (of the "Switches at" value) to cause a
switch from High to Normal Range.

High Range Values

High End

0.000% O

2

- 25.00% O

2

Low End

0.000% O

2

- 25.00% O

2

Enter the upper and lower analog output range values
for High Operating Range. The High End value defines
the measured O

2

value corresponding to the high

analog output value, i.e., 20mA or 10V, and the Low
End value corresponds to the low analog output value,
i.e., 0mA, 4mA, or 0V.

NOTE: Relay output can be initiated upon range change. (See page 5-12 of Table 5-5.)

Table 5-6. Efficiency Constants

UNITED STATES

EUROPE

CONSTANT

GAS

OIL

GAS

OIL

K1

0.407

0.432

0.66

0.69

K2

0.0

0.0

0.0082

0.0051

K3

5.12

5.12

12.28

8.74