Calibration 4-5

World Class 3000

37840018

CALIBRATION

GAS KIT #1

(P/N 6296A27G01)

Analytical

Figure 4-2. Portable Rosemount Analytical

Oxygen Calibration Gas Kit

2 Extra gas bottles are available

at:

Rosemount Analytical Inc.
6565 Davis Industrial Parkway
Solon, OH 44139-3922
U.S.A.

Rosemount Limited
Burymead Road
Hitchin, Herts. U.K.

Rosemount Italy
VIA Guido Cavalcanti 8
20127 Milan, Italy

Rosemount Spain
Saturnino Calleja 6
28002 Madrid
Spain

Rosemount France
165 Boulevard de Vallmy
92706, Colombes, France

Rosemount Analytical P/N
3530B07G01 for probe 0.4%
oxygen in nitrogen in dispos-
able bottle.

Rosemount Analytical P/N
3530B07G02 for probe 8%
oxygen in nitrogen in dispos-
able bottle.

3 A check valve is required at

the probe (between the cali-
bration fitting and the gas line)
to prevent the migration of
process gases down the cali-
bration gas line.

(b) Calibration with a Portable Rose-

mount Analytical Oxygen Calibra-
tion Gases Kit.

1 A typical portable calibration

setup is shown in Figure 4-3.
For manual (semiautomatic)
calibration, remove cap plug
from the calibrate in place fit-
ting. The cap plug must be
retained to seal this fitting af-
ter calibration is complete;
failure to do so may render
the probe useless if the sys-
tem pressure is slightly nega-
tive. The reference air should
be flowing as in normal
operation.

2 Refer to paragraph 4-2.d of

this section for Manual (Semi-
automatic) Calibration setup
and procedure using the IFT.

3 Screw the pushbutton regu-

lator with contents gage on to
the calibration gas of choice
and inject the calibration gas
by opening the valve. Gas is
on continuously when the
valve is opened.