World class 3000, Instruction manual – Emerson ROSEMOUNT 3000 User Manual

Page 44

Advertising
background image

Instruction Manual

IB-106-300NH Rev. 4.3
May 2005

2-8 Installation

Rosemount Analytical Inc. A Division of Emerson Process Management

World Class 3000

c. Reference Air Package

After the oxygen analyzing (probe) unit is
installed, connect the reference air set to
the probe junction box. The reference air
set should be installed in accordance with
Figure 2-3.

d. Service

Required

1. Power input: 100, 115 or 220 Vac sin-

gle phase, 50 to 60 Hz, 3 amp mini-
mum. (See label.)

2. Compressed air: 10 psig (68.95 kPa)

minimum, 225 psig (1551.38 kPa)

maximum at 2 scfh (56.6 L/hr) maxi-
mum; supplied by one of the following
(less than 40 parts-per-million total hy-
drocarbons). Regulator outlet pressure
should be set at 5 psi (35 kPa).

(a) Instrument air - clean, dry.

(b) Bottled standard air with step-down

regulator.

(c) Bottled compressed gas mixture

(20.95% oxygen in nitrogen).

(d) Other equivalent clean, dry, oil-free

air supply.

TO PROBE HEAD

REF AIR SET

263C152G01

1

FLOWMETER

0.2-2.0 SCFH 771B635H02

2

2" PRESSURE GAGE

0-15 PSIG

275431-006

3

COMBINATION FILTER-REG.

0-30 PSIG

4505C21G01

NOTE: DIMENSIONS ARE IN INCHES WITH

MILLIMETERS IN PARENTHESES.

1

2

3

4.81 (122.17)

FLOW SET

POINT KNOB

0.125-27 NPT FEMALE

OUTLET CONNECTION

1.19
(30.22)

10.0 REF

(254)

DRAIN VALVE

3.12 (79.25) MAX

8.50 MAX

(215.90)

2.0

(50.80)

2 MOUNTING HOLES

3.19 (81.03) LG

THROUGH BODY FOR

0.312 (7.92) DIA BOLTS

1.50

(38.10)

2.250 (57.15)

SCHEMATIC HOOKUP FOR REFERENCE AIR SUPPLY ON OXYGEN ANALYZER PROBE HEAD.

OUTLET

0.25-18 NPT FEMALE

INLET CONNECTION

COMPRESSED AIR SUPPLY
10-225 PSIG MAX PRESSURE

27270003

0.250 OR 6 MM OD TUBING

(SUPPLIED BY CUSTOMER)

0.250 OR 6 MM OD

TUBE COMPRESSION

FITTING (SUPPLIED BY WECO)

Figure 2-3. Air Set, Plant Air Connection