Tuning button: The tuning button fine-tunes the

mattress system's auto-adjusted pressure. Three tuning
settings are available:
0 System auto-adjusted pressure
- One level softer than auto-adjusted pressure
+ One level firmer than auto-adjusted pressure
Note: The tuning function is only available in alternate and
static therapy modes.

Front panel, second row

Therapy adjust button: The therapy ad-

just button selects the therapy mode. After power-up, the
system automatically inflates the mattress, auto-adjusts
the pressure, and enters alternate therapy mode. To then
cycle among the therapy settings, press the therapy adjust
button to select static therapy mode; press again to select
alternate + seat inflation therapy mode; press again to
select static + seat inflation therapy mode; press again to
select alternate therapy mode, etc.
Alternate: Each tube fills for 10 minutes, then deflates for
10 minutes, alternately to the next tube. When selected,
the alternate indicator illuminates.
Static: The pressure inside all tubes is adjusted to the
same pressure. When selected, the static indicator illumi-
nates.
Alternate + Seat Inflation: For use when the head is
raised, to increase pressure under the buttocks — in alter-
nating pressure therapy mode. When selected, both the
alternate and seat inflation indicators illuminate.
Static + Seat Inflation: For use when the head is raised, to
increase pressure under the buttocks — in static therapy
mode. When selected, both the static and seat inflation
indicators illuminate.