Graham Field ALTADYNE 765000 User Manual
Page 10
 
10
765000-INS-LAB-RevB10
Tuning button: The tuning button fine-tunes the
mattress system's auto-adjusted pressure. Three tuning 
settings are available:
0 System auto-adjusted pressure
- One level softer than auto-adjusted pressure
+	 One	level	firmer	than	auto-adjusted	pressure
Note: The tuning function is only available in alternate and 
static therapy modes.
Front panel, second row
Therapy adjust button: The therapy ad-
just	button	selects	the	therapy	mode.	After	power-up,	the	
system	automatically	inflates	the	mattress,	auto-adjusts	
the	pressure,	and	enters	alternate	therapy	mode.	To	then	
cycle	among	the	therapy	settings,	press	the	therapy	adjust	
button	to	select	static	therapy	mode;	press	again	to	select	
alternate	+	seat	inflation	therapy	mode;	press	again	to	
select	static	+	seat	inflation	therapy	mode;	press	again	to	
select	alternate	therapy	mode,	etc.
Alternate:	Each	tube	fills	for	10	minutes,	then	deflates	for	
10	minutes,	alternately	to	the	next	tube.	When	selected,	
the alternate indicator illuminates.
Static: The pressure inside all tubes is adjusted to the 
same	pressure.	When	selected,	the	static	indicator	illumi-
nates.
Alternate + Seat Inflation: For use when the head is 
raised,	to	increase	pressure	under	the	buttocks	—	in	alter-
nating	pressure	therapy	mode.	When	selected,	both	the	
alternate and seat inflation indicators illuminate.
Static + Seat Inflation:	For	use	when	the	head	is	raised,	to	
increase	pressure	under	the	buttocks	—	in	static	therapy	
mode.	When	selected,	both	the	static	and	seat	inflation	
indicators illuminate.