Graham Field ALTADYNE 765000 User Manual
Page 9
 
765000-INS-LAB-RevB10
9
PUMP: FRONT PANEL
Front panel of pump is shown below. Descriptions of but-
tons and indicators follow.
Tuning button;
indicators at left of button
Power failure indicator
Alarm mute button
Service indicator
Power (ON/OFF) button;
indicator at right of button
Panel lock button;
indicator at right of button
Maxfirm button
Auto-Adjust button;
indicators at right of button
Low pressure
indicator
Therapy adjust button;
indicators at right of button
Front panel, first row
Auto-Adjust button: The auto-adjust button
adjusts the mattress pressure automatically based on the 
patient's weight. To remind the user that the auto-adjust 
process	is	ongoing,	the	auto-adjust	indicator	LEDs	will	
periodically	flash	from	left	to	right	until	the	Ready	LED	illu-
minates to show the process is complete. Three conditions 
trigger the auto-adjust function:
1.	 Press	the	auto-adjust	button	for	two	seconds;	the	mat-
tress system automatically readjusts pressure setting.
2. When initial mattress inflation is complete, the mattress
system automatically starts the auto-adjust function.
3. When the mattress system detects a significant change
in patient's weight on the mattress (e.g. patient egress 
or	ingress),	the	system	automatically	starts	the	auto-
adjust function.
When	auto-adjust	is	complete,	the	Ready	indicator	will	illu-
minate.	To	disable	the	auto-adjust	function,	press	the	auto-
adjust button until auto-adjust indicators de-illuminate.
Note: During the auto-adjust operation, it is normal for the 
mattress system to cycle through a series of inflation and 
deflation.