TRANSPORT CAP
In	case	of	power	failure	or	trans-
port:	Disconnect	the	quick	connec-
tor and cap the mattress with the 
transport cap to slow deflation.
Note: This system is equipped 
with cell-on-cell technology to 
provide added support and prevent 
bottoming out in the event of a 
power failure. Lower cells will hold 
air for approximately eight hours.
CPR FUNCTION
If	CPR	must	be	performed	on	the	patient	while	the	mat-
tress	is	in	use,	to	immediately	deflate	the	mattress,	either:
1.	 Pull	the	CPR	valves	from	the	
mattress in the location of the 
arrow	shown	at	right,	or
2. Disconnect the quick connector
from the pump in the location of 
the arrow shown at right.
quick connector
transport cap