Graham Field ALTADYNE 765000 User Manual

Page 22

Advertising
background image

22

765000-INS-LAB-RevB10

Guidance and Manufacturer’s Declaration- Electromagnetic Immunity:
This device is intended for use in the electromagnetic environment specified below. The user of this

device should make sure it is used in such an environment.

Immunity Test

IEC60601 test

level

Compliance

Electromagnetic

Environment-Guidance

Conducted RF
IEC 61000-4-6

Radiated RF
IEC 61000-4-3

3Vrms150 kHz to
80 MHz outside
ISM bands

a

3 V/m 80 MHz to
2.5 GHz

3Vrms

3V/m

Portable

and

mobile

RF

communications

equipment

should be used no closer to any
part of this device, including
cables, than the recommended
separation distance calculated
from the equation applicable to
the frequency of the transmitter.

Recommended

separation

distance

P

d

2

.

1

=

150kHz to 80MHz

P

d

2

.

1

=

150kHz to 80MHz

P

d

3

.

2

=

80 MHz to 2.5G MHz

Where P is the maximum output
power rating of the transmitter in
watts (W) according to the
transmitter manufacturer and d is
the recommended separation
distance in meters (m).

b

Field strengths from fixed RF
transmitters, as determined by
an electromagnetic site survey

c

,

should be less than the
compliance level in each
frequency range

d

.

Interference may occur in the
vicinity of equipment marked with
the following symbol:

NOTE 1 At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by
absorption and reflection from structures, objects and people.