Graham Field ALTADYNE 765000 User Manual
Page 22
 
22
765000-INS-LAB-RevB10
Guidance and Manufacturer’s Declaration- Electromagnetic Immunity: 
This device is intended for use in the electromagnetic environment specified below. The user of this 
device should make sure it is used in such an environment.
Immunity Test
IEC60601 test
level
Compliance
Electromagnetic
Environment-Guidance
Conducted RF 
IEC 61000-4-6 
Radiated RF 
IEC 61000-4-3 
3Vrms150 kHz to 
80 MHz outside 
ISM bands
a
3 V/m 80 MHz to 
2.5 GHz 
3Vrms
3V/m
Portable
and
mobile
RF
communications
equipment
should be used no closer to any 
part of this device, including 
cables, than the recommended 
separation distance calculated 
from the equation applicable to 
the frequency of the transmitter. 
Recommended
separation
distance
P
d
2
.
1
=
150kHz to 80MHz
P
d
2
.
1
=
150kHz to 80MHz
P
d
3
.
2
=
80 MHz to 2.5G MHz
Where P is the maximum output 
power rating of the transmitter in 
watts (W) according to the 
transmitter manufacturer and d is 
the recommended separation 
distance in meters (m).
b
Field strengths from fixed RF 
transmitters, as determined by 
an electromagnetic site survey
c
,
should be less than the 
compliance level in each 
frequency range
d
.
Interference may occur in the 
vicinity of equipment marked with 
the following symbol: 
NOTE 1 At 80 MHz and 800 MHz, the higher frequency range applies. 
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by 
absorption and reflection from structures, objects and people.