Low pressure indicator: The low pressure
indicator	(yellow	LED)	illuminates,	and	the	alarm	sounds,	
when	an	abnormally	low	pressure	level	occurs.	If	the	low	
pressure indicator persistently illuminates inappropriately:
1.	 Check	to	ensure	all	connections	are	properly	and	se-
curely connected per installation instructions.
2. Check for any leakage (tubes or connecting hoses). If
necessary,	contact	your	Graham-Field	equipment	pro-
vider to replace any damaged tubes or hoses.
3. If problem persists, contact your Graham-Field equip-
ment provider.
Note: Sometimes the low pressure indicator will illuminate 
when a patient gets up to leave the bed. This is normal. The 
indicator will turn off automatically after a short time.
The low pressure indicator will illuminate until the low-
pressure condition is resolved.
Power failure indicator: The power failure
indicator	illuminates,	and	the	alarm	sounds,	when	there	is	
a power failure. Depress the alarm mute button to disable 
both	alarm	and	LED.	
Service indicator: Mechanical failure, service
required; contact your Graham-field equipment provider.