APPENDIX A: EMC INFORMATION
Guidance and Manufacturer’s Declaration- Electromagnetic Emissions:
This device is intended for use in the electromagnetic environment specified below. The user of this device should 
make sure it is used in such an environment. 
Emissions Test
Compliance
Electromagnetic Environment-Guidance
Harmonic emissions 
IEC61000-3-2 
Class A
Voltage fluctuations / 
Flicker emissions 
IEC61000-3-3 
Complies
The device is suitable for use in all establishments, including 
domestic establishments and those directly connected to the 
public low-voltage power supply network. 
Guidance and Manufacturer’s Declaration- Electromagnetic Immunity:
This device is intended for use in the electromagnetic environment specified below. The
user of this device should make sure it is used in such an environment.
Immunity Test IEC60601 test level
Compliance
Electromagnetic
Environment-Guidance
Electrostatic 
Discharge(ESD)
IEC61000-4-2 
±6kV contact 
±8kV air 
±6kV contact 
±8kV air 
Floors should be wood, concrete or 
ceramic tile. If floors are covered 
with synthetic material, the relative 
humidity should be at least 30 %. 
Electrical fast 
transient/ burst 
IEC61000-4-4 
±2kV for power supply 
line
±1kV for input/out line 
±2kV for power supply 
line
±1kV for input/out line 
Mains power quality should be that 
of atypical commercial or hospital 
environment
Surge
IEC61000-4-5 
±1kV for differential 
mode
±2kV for common mode
±1kV for differential 
mode
±2kV for common mode
Mains power quality should be that 
of atypical commercial or hospital 
environment.
Voltage dips, 
short
interruptions and 
voltage 
variations on 
power supply 
input lines 
IEC61000-4-11 
<5 % UT(>95 % dip in 
UT)for 0,5 cycle 
40 % UT(60 % dip in 
UT)for 5 cycles 
70 % UT(30 % dip in 
UT)for 25 cycles 
<5 % UT(>95 % dip in 
UT)for 5 sec 
<5 % UT(>95 % dip in 
UT) for 0,5 cycle 
40 % UT(60 % dip in 
UT) for 5 cycles 
70 % UT(30 % dip in 
UT) for 25 cycles 
<5 % UT(>95 % dip in 
UT) for 5 sec 
Mains power quality should be that 
of atypical commercial or hospital 
environment. If the user of this 
device requires continued operation 
during power mains interruptions, it 
is recommended that the device be 
powered from an uninterruptible 
power supply or a battery. 
Power frequency 
(50/60Hz) 
magnetic field 
IEC61000-4-8 
3 A/m
3 A/m
Power frequency magnetic fields 
should be at levels characteristic of 
atypical location in a typical 
commercial or hospital environment.
NOTE: U
T
is the a.c. mains voltage prior to the application of the test level