— how many tests have been run. Four figures are shown: daily

(upper left), weekly (upper right), and monthly (lower left) periodic
selftests, and battery insertion selftests (lower right).

REV

— device language, model, and software revision.

Battery History

Information about use of the battery currently installed in your FR2+ is also
available. To review the history of the battery, select

NEXT

from the menu

screen displayed when you insert the battery, then select

BATTERY

HISTORY

from the next menu displayed.

The battery history information is read from the internal memory of the battery.
It includes:

USE

MINUTES

— the total operating time (in minutes), including selftest

time, for this battery;

CHARGES

— the total number of full defibrillation charges that have been

provided by this battery, including selftest charges;

BATTERY

— a

GOOD

BATTERY

(M3863A) or a fuel gauge display

(M3848A) showing 25%, 50%, 75% or 100%, or a

LOW

BATTERY

or

REPLACE

BATTERY

message, as appropriate.

STATUS

— the current status of this battery, displayed in a binary code.

Make a note of this code if technical service is needed.