S e t u p a n d A d v a n c e d M o d e F e a t u r e s

6-3

P

h

ili

ps

Me

dic

a

l

Sy

s

tem

s

6

NSA action

(minutes)

MONITOR

,

0.5, 1.0, 1.5,

2.0, 2.5, 3.0

MONITOR

Sets how the FR2+ behaves following a

NO

SHOCK

ADVISED

(NSA) decision:

MONITOR

— directs the FR2+ to monitor the patient’s

ECG following an NSA decision and to prompt the user
periodically to provide CPR. The interval for CPR
prompting is set by the Monitor Prompt Interval.

TIME

SETTING

— directs the FR2+ to provide a CPR

pause period following an NSA decision (NSA CPR
Pause).

• If no shocks have been delivered in the current Shock

Series (e.g., the patient’s initial monitored rhythm is
non-shockable), the length of the CPR pause is defined
by the NSA Action time setting.

• If shocks have been delivered in the current Shock

Series (e.g., the NSA decision follows a shock), the
length of the CPR pause is instead defined by the CPR
Timer
setting.

CPR prompt

LONG

,

SHORT

LONG

Sets the level of detail provided in the CPR reminder voice
prompts provided at the completion of a Shock Series.

LONG

— provides detailed coaching to check airway,

breathing, and pulse before beginning CPR.

SHORT

— simply directs user to begin CPR if needed.

monitor prompt
interval
(minutes)

1.0, 1.5, 2.0,

2.5, 3.0,

(infinite)

1.0

Sets the interval for patient care prompts provided during
FR2+ monitoring of the patient’s ECG following an NSA
decision. Selection of

(infinite) means that no repeat

prompting will be provided during ECG monitoring.

parameter

settings

default

description