Sending and receiving clock settings – Philips M3860A User Manual

Page 56

Advertising
background image

M3860A/M3861A HEARTSTART FR2+

6-8

P

h

ili

ps

Me

dic

a

l

Sy

s

tem

s

“receiving” FR2+ displays a

SETUP

COMPLETE

message. Your

HeartStart FR2+ immediately uses the new setup.

Receiving setup from a computer running HeartStart Event Review software
is discussed in the directions for use provided with HeartStart Event Review
software.

Reading setup

This method copies setup data from a data card to your HeartStart FR2+. To
read the setup, follow these steps:

1.

Insert the data card in the data card tray and install the loaded tray into
the data card slot in the FR2+.

2.

Select

READ

SETUP

from the setup menu.

3.

A new screen comes up. If the FR2+ cannot read the data card or
cannot find a valid setup on the data card, the screen displays a

NO

SETUP

FILE

error message. Otherwise, the FR2+ begins reading the

setup information from the data card immediately.

4.

If you select

EXIT

before the transfer is complete, the revised setup will

not be copied. When the transfer is finished, the screen displays a

SETUP

COMPLETE

message. Your FR2+ immediately uses the revised

setup.

Sending and Receiving Clock Settings

To synchronize the clock settings of your HeartStart FR2+ with the clock of
another FR2+ or a computer running HeartStart Event Review software, you
can use the infrared communications feature.

Instructions for synchronizing clock settings using a computer running
HeartStart Event Review are provided with the HeartStart Event Review
software.

To transfer clock settings from one FR2+ to another:

1.

Remove and reinstall the battery of both FR2+ devices to bring up the
first menu screen.

2.

Select

NEXT

to go to the second menu screen.

3.

Select

CLOCK

from the second menu screen. The

CLOCK

screen then

comes up.

4.

Locate the infrared communications port on each FR2+ and line them up
with one another, so that the infrared “eye” in each one has an

Advertising