Philips M3860A User Manual

Page 76

Advertising
background image

M3860A/M3861A HEARTSTART FR2+

B-8

P

h

ili

ps

Me

dic

a

l

Sy

s

tem

s

(Optional) M3854A data card

(Optional) M3864A training & administration pack

(Optional) M3855A charger

Defibrillator Pad
Requirements

Use only DP2/DP6, M3870A, M3713A, and M3716A defibrillator pads with the
HeartStart FR2+. Place the pads on the patient as illustrated on each pad.

category

nominal specifications

category

nominal specifications

Capacity

8 hours of event and ECG data, or 60 minutes with voice recording.

category

nominal specifications

Battery Type

12 V, 1.1 Ah, nickel metal hydride. Disposable, rechargeable cell using the
M3855A charger.

Capacity

Provides 4 hours of operating time at 77 °F (25 °C).

Status Indicators

Low battery: flashing red X on the front panel of the FR2+.
Dead battery: solid red X on the front panel of the FR2+.

Storage Temperature

50° to 104° F (10° to 40° C).

category

nominal specifications

Application

For use with M3864A training & administration pack only.

Power Requirements

With appropriate power cord, any AC mains power input or inverter-type power
sources.

Environmental
Requirements

32° to 113° F (0° to 45° C).
35% to 85% relative humidity (non-condensing).

Conformance Testing

International: EN60335-1:1994 Class I
North America: UL 1310 Class 2