Philips M3860A User Manual

Page 77

Advertising
background image

Te c h n i c a l S p e c i f i c a t i o n s

B-9

P

h

ili

ps

Me

dic

a

l

Sy

s

tem

s

B

(Optional) M3873A/M3874A FR2+ ECG assessment
module

category

nominal specifications

Application

For use with the FR2+ M3860A with ECG display enabled and running version
1.5 software or higher (denoted by FR2+ on the front panel or rear label).

Length and Weight

100 inches (182 cm);

≤ 1 lb. (2.2 kg).

Operating Temperature and
Humidity

32° to 122° F (0° to 50° C);
0% to 95% relative humidity (non-condensing).

Storage Temperature and
Humidity

32° to 109° F (0° to 43° C).
0% to 75% relative humidity (non-condensing).

Patient Lead Wire
Designation

M3873A (AAMI):

M3874A (IEC):

positive lead

− red

positive lead

− green

negative lead

− white

negative lead

− red

reference lead

− black.

reference lead

− yellow

Typical (Lead II) Connection

Lead II vectors:

positive

− left leg

negative

− right arm

reference

− left arm.

Other limb vectors can be obtained by different electrode positions.

Battery Type

3 V, 1 Ah, poly-carbonmonofluoride lithium (LiCFx). Non-replaceable disposable
primary cell.

Service Life

Typically, 5 years.

Performance with FR2+
Defibrillator

Meets environmental specifications cited for FR2+ Defibrillator on page B1
through B2.