Defibrillation warnings and cautions

CAUTION

Use only Philips-approved data cards. The HeartStart FR2+ may perform
improperly if non-approved accessories are used.

safety level

possible electrical interference with ECG monitoring

WARNING

Radio-frequency (RF) interference from devices such as cellular
phones and two-way radios can cause improper HeartStart FR2+
operation. The HeartStart FR2+ should be used at least 6 feet (2
meters) away from RF devices, as stated in accordance with EN
61000-4-3:1996.

safety level

possible improper device performance

safety level

possible shock hazard

WARNING

Defibrillation current can cause operator or bystander injury. Do not
touch the patient during defibrillation. Disconnect the pads connector
from the HeartStart FR2+ before using any other defibrillator.

safety levels

possible ECG misinterpretation

WARNING

For safety reasons, some very low-amplitude or low-frequency heart
rhythms may not be interpreted by the HeartStart FR2+ as shockable
VF rhythms. Also, some VT rhythms may not be interpreted as
shockable rhythms.

WARNING

Handling or transporting the patient during heart rhythm analysis can
cause an incorrect or delayed diagnosis. If the HeartStart FR2+ gives
a

SHOCK

ADVISED

prompt, keep the patient as still as possible for at

least 15 seconds so the HeartStart FR2+ can reconfirm the rhythm
analysis before a shock is delivered.