Te c h n i c a l S p e c i f i c a t i o n s

B-3

P

h

ili

ps

Me

dic

a

l

Sy

s

tem

s

B

Energy

Using adult defibrillator pads: 150 J nominal into a 50 ohm load.
Using infant/child reduced-energy defibrillator pads: 50 J nominal into a 50 ohm
load. Sample pediatric energy doses:

age

energy dose

newborn

14 J/kg

1 year

5 J/kg

2

− 3 years

4 J/kg

4

− 5 years

3 J/kg

6

− 8 years

2 J/kg

Doses indicated are based on CDC growth charts for the 50th percentile weights
for boys.*

*

National Center for Health Statistics in collaboration with the National Center for
Chronic Disease Prevention and Health Promotion. CDC growth charts:
weight-for-age percentiles, revised and corrected
November 28, 2000. Atlanta,
GA: Centers for Disease Control and Prevention © 2000.

Charge Control

Controlled by Patient Analysis System for automated operation. Can be
programmed for manual initiation using advanced mode of the M3860A.

Charge Time
from “Shock Advised”

< 10 seconds typical, including confirming analysis. Charge time increases near
end of battery service life.

Shock-to-Shock Cycle Time

< 20 seconds typical, including analysis, in AED mode.

“Charge Complete”
Indicator

Shock button flashes, audio tone sounds.

Disarm (AED mode)

Once charged, the HeartStart FR2+ will disarm if:
• patient’s heart rhythm changes to non-shockable rhythm, OR
• a shock is not delivered within 30 seconds after the FR2+ is armed, OR
• the PAUSE button (if enabled) is pressed, OR
• the On/Off button is pressed to turn off the FR2+, OR
• the defibrillator pads are removed from the patient or the pads connector is

disconnected from the FR2+.

category

nominal specifications