Figure 1-6. clk100 termination – National Instruments NI PXIe-1075 User Manual

Page 18

Advertising
background image

Chapter 1

Getting Started

© National Instruments Corporation

1-11

NI PXIe-1075 User Manual

An independent buffer drives PXIe_CLK100 to each peripheral slot. These
clocks are matched in skew to less than 100 ps. The differential pair must
be terminated on the peripheral with LVPECL termination for the buffer to
drive PXIe_CLK100 so that when there is no peripheral or a peripheral that
does not connect to PXIe_CLK100, there is no clock being driven on the
pair to that slot. Refer to Figure 1-6 for a termination example.

Figure 1-6. CLK100 Termination

An independent buffer drives PXIe_SYNC100 to each peripheral slot. The
differential pair must be terminated on the peripheral with LVPECL
termination for the buffer to drive PXIe_SYNC100 so that when there is
no peripheral or a peripheral that does not connect to PXIe_SYNC100,
there is no SYNC100 signal being driven on the pair to that slot. Refer to
Figure 1-6 for a termination example.

In summary, PXI_CLK10 is driven to every slot. PXIE_CLK100 and
PXIE_SYNC100 are driven to every peripheral slot.

PXI_CLK10, PXIe_CLK100 and PXIe_SYNC100 have the default timing
relationship described in Figure 1-7.

Figure 1-7. System Reference Clock Default Behavior

CLK100 +

CLK100 –

50

Ω

50

Ω

47

Ω

0.01 µF

+

PXIe_CLK100

PXI_CLK10

PXIe_SYNC100

0 1 2 3 4 5 6 7 8 9 0 1 2 3 4 5 6 7 8 9 0 1 2 3 4 5 6 7 8 9

Advertising