Operating Environment

Ambient temperature range.................... 0 to 55 °C

(Tested in accordance with
IEC-60068-2-1 and
IEC-60068-2-2. Meets
MIL-PRF-28800F Class 3
low temperature limit and
MIL-PRF-28800F Class 2
high temperature limit.)

Relative humidity range ......................... 10 to 90%, noncondensing

(Tested in accordance with
IEC-60068-2-56.)

Storage Environment

Ambient temperature range.................... –40 to 71 °C

(Tested in accordance with
IEC-60068-2-1 and
IEC-60068-2-2. Meets
MIL-PRF-28800F Class 3 limits.)

Relative humidity range ......................... 5 to 95%, noncondensing

(Tested in accordance with
IEC-60068-2-56.)

Shock and Vibration

Operational shock .................................. 30 g peak, half-sine, 11 ms pulse

(Tested in accordance with
IEC-60068-2-27. Meets
MIL-PRF-28800F Class 2 limits.)

Random Vibration.................................. 5 to 500 Hz, 0.3 g

rms

Acoustic Emissions

Sound Pressure Level (at Operator Position)

(Tested in accordance with ISO 7779. Meets MIL-PRF-28800F
requirements.)

Auto fan (up to ~30 °C ambient) ........... 45.0 dBA

High fan.................................................. 63.3 dBA