Chassis cooling – National Instruments NI PXI-1056 User Manual

Page 39

Advertising
background image

Appendix A

Specifications

© National Instruments Corporation

A-3

NI PXI-1056 User Manual

Over-current protection.......................... All outputs protected from short

circuit and overload with
automatic recovery when the
short or overload is removed.

Over-voltage protection

1

3.3 V, 5 V, +12 V, –12 V................ 130% above nominal output

voltage

Power-supply shuttle MTTR.................. Replacement in under 5 minutes

Chassis Cooling

Per slot cooling capacity ........................ Slot cooling capacity is 25 W

with fan speed set to HIGH

Slot airflow direction ............................. P1 to P2, bottom of module

to top of module

Module cooling

System............................................. Forced air circulation (positive

pressurization) through one
220 cfm fan with HIGH/AUTO
speed selector

Intake .............................................. Right rear of chassis

Exhaust............................................ Along left side of chassis

Power supply cooling

System............................................. Forced air circulation through

integrated fan

Intake .............................................. Right rear of chassis

Exhaust............................................ Along left side of chassis

1

To reset over-voltage protection, remove the mains for 30 seconds and then power back up.