Chapter 12 triggering, Triggering with a digital source, Figure 12-1. falling-edge trigger – National Instruments NI 6238 User Manual

Page 131: Triggering with a digital source -1, Triggering

Advertising
background image

© National Instruments Corporation

12-1

NI 6238/6239 User Manual

12

Triggering

A trigger is a signal that causes an action, such as starting or stopping the
acquisition of data. When you configure a trigger, you must decide how you
want to produce the trigger and the action you want the trigger to cause.
NI 6238/6239 devices support internal software triggering, as well as
external digital triggering. For information about the different actions
triggers can perform for each sub-system of the device, refer to the
following sections:

The

Analog Input Triggering

section of Chapter 4,

Analog Input

The

Analog Output Triggering

section of Chapter 5,

Analog Output

The

Counter Triggering

section of Chapter 7,

Counters

Triggering with a Digital Source

Your DAQ device can generate a trigger on a digital signal. You must
specify a source and an edge. The digital source can be any of the PFI,
RTSI, or PXI_STAR signals.

The edge can be either the rising edge or falling edge of the digital signal.
A rising edge is a transition from a low logic level to a high logic level. A
falling edge is a high to low transition.

Figure 12-1 shows a falling-edge trigger.

Figure 12-1. Falling-Edge Trigger

High

Falling Edge Initiates Acquisition

0 V

Digital Trigger

Advertising