Figure 4-5. ni 6238/6239 pgia, Table 4-2. signals routed to the ni-pgia – National Instruments NI 6238 User Manual

Page 32

Advertising
background image

Chapter 4

Analog Input

NI 6238/6239 User Manual

4-6

ni.com

Figure 4-5. NI 6238/6239 PGIA

Analog input ground-reference setting refers to the reference that the PGIA
measures against. Differential is the only ground-reference setting for NI
6238/6239 analog input signals, which means that the PGIA always
measures the voltages between AI + and AI – generated across the input
sense resistor, regardless of which analog input connection method is used.
Refer to the

Connecting Analog Current Input Signals

section for allowable

input connection methods.

When measuring voltages internal to the device, such as the onboard
reference, the signal is measured against AI GND. This measurement
method is called referenced single-ended (RSE). The name is derived from
the fact that one end of the measurement, the positive input of the PGIA, is
connected to a signal, and the negative input of the PGIA is connected to
the AI GND reference.

Table 4-2 shows how signals are routed to the NI-PGIA.

Note On devices that allow multiple ground-reference settings, the settings are
programmed on a per-channel basis.

Table 4-2. Signals Routed to the NI-PGIA

AI Ground-Reference

Settings

Signals Routed to the Positive

Input of the NI-PGIA (V

in+

)

Signals Routed to the Negative

Input of the NI-PGIA (V

in–

)

DIFF

AI <0..7>+

AI <0..7>–

RSE

Internal Channels

AI GND

I

in+

V

m

= I

in

R

× Gain

V

m

I

in–

PGIA

+

Measured

Voltage

Instrumentation

Amplifier

Current

Sense

Resistor

AI GND

Advertising
This manual is related to the following products: