M30880T-EPB User’s Manual

2. Setup

REJ10J0497-0100Z Rev.1.00 January 16, 2005

Page 27 of 100

2.8 Self-check


2.8.1 Self-check Procedure
To run the emulator self-check, do so as explained here below. While the self-check is in progress, LEDs will change as shown
in Figure 2.6.

(1) If the user system is connected, disconnect it.
(2) Set the switches as the factory-settings to execute the self-check (see Table 2.1).
(3) Within 2 seconds of activating power to the emulator, press the system reset switch on the emulator front panel to switch

the emulator to maintenance mode.

(4) Check the "SAFE" LED starts flashing and then press the system reset switch again.
(5) The self-check will start. If the normal result is displayed in about 20 seconds, the self-check terminated normally.

Table 2.1 Switch settings at self-check

Switch Setting

AVCC/AVSS switch (SW1)

EXT side

P87 switch (SW2)

P87 side

P86 switch (SW3)

P86 side

XOUT switch (SW4)

NC side

A/D conversion bit selection switch (SW5)

All upper side

: ON
: OFF
: Flashing

POW
ER
CLOCK

RESET

RUN

WARNING

POWER

SAFE

ERROR

SYSTEM

STATUS

TARGET

STATUS

Maintenance mode started

POWER

CLOCK

RESET

RUN

WARNING

POWER

SAFE

ERROR

SYSTEM

STATUS

TARGET

STATUS

Self-check terminated normally

POWER

CLOCK

RESET

RUN

WARNING

POWER

SAFE

ERROR

SYSTEM

STATUS

TARGET

STATUS

Emulation probe and firmware

do not match

・When the firmware does not support the emulation probe
・When the emulation probe is not connected properly

POWER

CLOCK

RESET

RUN

WARNING

POWER

SAFE

ERROR

SYSTEM

STATUS

TARGET

STATUS

PC7501 system status error

・When the PC7501 system is not working properly

POWER

CLOCK

RESET

RUN

WARNING

POWER

SAFE

ERROR

SYSTEM

STATUS

TARGET

STATUS

Emulation probe initialization error

・When a clock is not supplied to the emulation probe

POWER

CLOCK

RESET

RUN

WARNING

POWER

SAFE

ERROR

SYSTEM

STATUS

TARGET

STATUS

Emulation probe status error

・When the emulation probe is not working properly

ERROR 1

SAFE and ERROR flash

ERROR 2

SAFE flashes and

ERROR lights

ERROR 3

SAFE and ERROR

light alternately

ERROR 4

SAFE goes out and

ERROR lights

Figure 2.6 Self-check procedure