Calibration

Exit calibration mode

Release 2.1

129

Exit calibration mode

When you have completed calibration, exit calibration mode by sending the
command:

CAL:STAT OFF, "0000"

where “0000” is replaced with your security code.

Restore Factory Calibration

Restore the unit to the calibration constants set at the factory by sending the
command:

CALibration:RESTore

Note This procedure should not be used in place of regular calibration, but may be useful to

restore the unit to an operational state in case of failure.