Power On

Output State

(XFR/XHR only) You may also change the output state whether the output is enabled
or disabled at power on.

To change the power on at output state:
OUTP:PON:STAT [ON|OFF|1|0]

Whether the unit powers on with output enabled is a function of both the state at
power off and this setting.

Reset Resetting the unit puts certain features to a known state. These states are listed in

Table 3.2

.

To reset the unit:
*RST or SYST:RES

Table 3.2 Features Affected by Reset (*RST) Command

Feature

Reset State (*RST)

Voltage setpoint

0.0V

Current setpoint

0.0A

Low voltage setpoint limit

0.0V

High voltage setpoint limit

101% of voltage rating

Low current setpoint limit

0.0A

High current setpoint limit

101% of current rating

Over voltage protection

Disabled (0.0V)

Under voltage protection

Disabled (0.0V) and not shutdown when tripped

Over current protection

Disabled (0.0A) and not shutdown when tripped

Under current protection

Disabled (0.0A) and not shutdown when tripped

Fold shutdown protection

None and delay 0.5s

AC off shutdown recovery

Auto recover

OTP shutdown recovery

Latched

Triggered voltage setpoint

Disabled (Default)

Triggered current setpoint

Disabled (Default)

Autosequence mode

Exit

Calibration mode

False

Output

Disabled