Table 4.9, Current sub-register, Figure 4.3 – Xantrex Technology GPIB-M-XT User Manual

Page 79: Ieee 488.2 status register and status byte, Table

Advertising
background image

Status Registers

Overview

Release 2.1

77

CURRent

Sub-Register

This shows whether the present current level is over or under the specified trip limit.

Table 4.9 CURRent Sub-Register

Standard

Event Status

Register

The standard event status register sets bits for specific events during power supply
operation. All bits in the standard event status registers are set through the error event
queue. The register is defined by 488.2 and is controlled using 488.2 common
commands, *ESE, *ESE?, and *ESR?.

Figure 4.3

summarizes the standard Event Status Register

Figure 4.3 IEEE 488.2 Status Register and Status Byte

Bit

Bit Weight

Bit Name

Description

0

1

Over CURrent

Set if the supply’s output current is
greater than the user-specified
over-current trip level (variable trip limit)
and the supply is in Operation state.

1

2

Under CURrent

Set if the supply’s output current is less
than the user-specified under current
trip level (variable trip limit) and the
supply is in Operation state.

Not Used
Not Used
Not Used

MAV

Standard Event Status Register

1

0

3

2

5

4

7

6

Summary of QUESTionable Status

RQS/MSS

Summary of OPERation Status

1

0

3

2

5

4

7

6

9

8

11

10

13

12

15

14

Reserved
Reserved

Reserved
Reserved

Reserved
Reserved

Reserved
Reserved

+

Operation Complete

Query Error

Device Dependent Error

User Request

Power On

Execution Error

Command Error

Not Used

Status Byte