Remote interface reference – American Magnetics 420 Power Supply Programmer User Manual

Page 104

Advertising
background image

86

Rev. 7

Remote Interface Reference

Ramp Configuration Commands and Queries

‡

CONFigure:RAMP:RATE:CURRent <

rate (A/s, A/min)

>

Sets the ramp rate in units of amperes/second or amperes/minute, per the
selected ramp rate units.

‡

RAMP:RATE:CURRent?

Returns the ramp rate setting in units of amperes/second or amperes/
minute, per the selected ramp rate units.

‡

CONFigure:RAMP:RATE:FIELD <

rate (kG/s, kG/min, T/s, T/min)

>

Sets the ramp rate in units of kilogauss/second or minute, or Tesla/second
or minute (per the selected field units and ramp units). This command
requires that a coil constant be defined, otherwise an error is generated.

‡

RAMP:RATE:FIELD?

Returns the ramp rate setting in units of kilogauss/second or minute, or
Tesla/second or minute (per the selected field units and ramp units). This
query requires that a coil constant be defined, otherwise an error is
generated.

‡

CONFigure:RAMP:CURRent <

current

>,<

rate (A/s, A/min)

>

Sets both the programmed current in amperes, and the ramp rate in
amperes/second or amperes/minute (per the selected ramp units), with one
command string. Note that both parameters are required.

‡

RAMP:CURRent?

Returns the programmed current in units of amperes and the ramp rate in
units of amperes/second or amperes/minute (per the selected ramp units).
The two return values are separated by a comma. For example:

RAMP:CURRENT?

50.0000,0.1000

‡

CONFigure:RAMP:FIELD

<

field (kG, T)

>,<

rate (kG/s, kG/min, T/s, T/min)

>

Sets both the programmed field in kilogauss or Tesla (per the selected field
units), and the ramp rate in kilogauss/second, kilogauss/minute, Tesla/
second, or Tesla/minute (per the selected field units and ramp units), with
one command string. Note that both parameters are required. This
command requires that a coil constant be defined, otherwise an error is
generated.

‡

RAMP:FIELD?

Returns the programmed field in units of kilogauss or Tesla (per the
selected field units) and the ramp rate in units of kilogauss/second,
kilogauss/minute, Tesla/second, or Tesla/minute (per the selected field
units and ramp units). The two return values are separated by a comma.