Example Setup

If your magnet, programmer, and power supply were purchased as a
system from AMI, the setup menus are preset by AMI to match the
magnet purchased.

Table 3-4 provides a summary of the Model 420 setup parameters for this
example.

Table 3-4. Example Setup Configuration

Select Power Supply

AMI 12100PS

Stability Setting (%)

0.0

Coil Constant

a

(kG/A)

a. Also referred to as the Field-to-Current Ratio. Obtained by

dividing the rated field by the rated current if not explicitly
stated.

1.1806

Current Limit (A)

76.300

PSwitch Installed?

YES

PSwitch Current (mA)

46.0

PSwitch Heated Time (sec)

15

Enable Quench Detect?

YES

Energy Absorber Present?

NO

Programmed Current (A)

76.2300

Voltage Limit (V)

4.000

b

b. Value is the 2 V charge rate plus allowances for power lead

drop at the rated current. If a Model 601 energy absorber is
present, add an additional 5 V to the value.

Ramp Rate (A/sec)

0.2041

c

c. Value is obtained by dividing the charge rate (V) by the

inductance (H).