41

Operation

Setup Menu : Supply

Note

The operating current must be less than 0.1% of I

max

in order to

change the supply selection. If a change is attempted above this
current, the Model 420 will beep and ignore the keypress. Power
supply selection should also preferably be performed with the power
supply off for maximum safety.

The power supply settings define the V-I ranges for a specific
supply. For example, V-I diagrams are presented in Figure 3-3 for
the AMI 12100PS and AMI 4Q05100PS selections. The AMI
12100PS operates as a one-quadrant system without the addition of
an energy absorber. With the addition of an energy absorber, the
AMI 12100PS system can function as a two-quadrant supply. The
AMI 4Q05100PS power supply operates as a four-quadrant power
supply without the addition of an energy absorber.

The addition of an energy absorber to the system does not change
the capabilities of the power supply (or the values entered for the
supply). The addition of an energy absorber does, however, change
the system operating ranges per the example of Figure 3-3.

Table 3-2. Available Select Power Supply options.

Power Supply

Min Output

Voltage (V)

Max Output

Voltage (V)

Min

Output

Current (A)

Max Output

Current (A)

V-V Mode Input

Range (V)

AMI 12100PS

+0.000

+12.000

+0.000

+100.000

+0.000 to +10.000

AMI 12200PS

+0.000

+12.000

+0.000

+200.000

+0.000 to +10.000

AMI 4Q05100PS

5.000

+5.000

100.000

+100.000

10.000 to +10.000

AMI 10100PS

+0.000

+10.000

+0.000

+100.000

+0.000 to +5.000

AMI 10200PS

+0.000

+10.000

+0.000

+200.000

+0.000 to +5.000

HP 6260B

+0.000

+10.000

+0.000

+100.000

+0.000 to +10.000

Kepco BOP 20-5M

a

10.000

+10.000

5.000

+5.000

10.000 to +10.000

Kepco BOP 20-10M

10.000

+10.000

10.000

+10.000

10.000 to +10.000

Xantrex XFR 7.5-140

+0.000

+7.500

+0.000

+140.000

+0.000 to +10.000

Custom...

b

20.000

+20.000

200.000

+200.000

10.000 to +10.000

a. The Kepco supplies are limited to only 1/2 of the output voltage range since the

supplies are only designed to safely dissipate 1/2 of the rated power output.

b. The values shown for the Custom... option are defaults. The user should enter the

appropriate values within the respective submenus. Custom values, once entered,
are saved between power-ups.