6 switch heater commands and queries, Switch heater commands and queries, Table 4-5 – American Magnetics 420 Power Supply Programmer User Manual

Page 106: Return values and their meanings for the, Remote interface reference

Advertising
background image

88

Rev. 7

Remote Interface Reference

Switch Heater Commands and Queries

‡

ZERO

Places the Model 420 in ZEROING CURRENT mode. Ramping
automatically initiates and continues at the ramp rate until the power
supply output current is equal to 0 A at which point the AT ZERO status is
returned.

‡

STATE?

Returns an integer value corresponding to the ramping state according to
the table below:

4.5.6

Switch Heater Commands and Queries

The

PSwitch

commands control and query the state of the persistent

switch heater. For further information regarding the persistent switch
heater, see paragraph 3.4 on page 59.

‡

PSwitch {0|1}

Turns the persistent switch heater OFF and ON. Sending “0” turns the
switch heater OFF. Sending a “1” turns the switch heater ON. The default
value is “0”.

‡

PSwitch?

Returns a “0” indicating the switch heater is OFF, or a “1” indicating the
persistent switch heater is ON.

Table 4-5. Return values and their meanings for the

STATE?

query.

Return

Value

Meaning

1

RAMPING to programmed current/field

2

HOLDING at the programmed current/field

3

PAUSED

4

Ramping in MANUAL UP mode

5

Ramping in MANUAL DOWN mode

6

ZEROING CURRENT (in progress)

7

Quench detected

8

Heating persistent switch

9

AT ZERO current