Introduction

Operating Characteristics

reversing the contacts and resuming magnet energization. This pause
precludes smooth magnetic field reversals.

1.5.4

True Four-Quadrant Operation

The true four-quadrant magnet power supply system illustrated in Figure
1-5 offers the most control of all the modes of operation. Efficiency is
increased and reversible magnetic field profiles are attainable without
discontinuities. All of the current switching is performed electronically so
that system reliability is improved. Disadvantages of the four-quadrant
system include the increased cost of the power supply if smooth,
continuous field polarity reversal is not a requirement, and added
complexity in protecting the power supply in the event of AC power loss or
quenches.

Magnet

Coil(s)

Persistent

Switch

(optional)

Misc. Line Losses

Model 420

Shunt

V

Four-Quadrant

Power Supply

Current

Figure 1-5. True Four-Quadrant System