Summary of Operational Limits

3.6 Summary of Operational Limits and Default Settings

Table 3-6 provides a summary of the operational limits and the default
setting for all Model 420 parameters. If the operator attempts to enter a
value outside of the limits, the Model 420 will beep once and revert to the
previous setting.

References to the specifications on page 7 indicate that the absolute limit
is determined by the specific configuration of the Model 420 purchased.

Table 3-6. Summary of limits and defaults for the Model 420.

Model 420 Setting (Units)

Absolute Limits

Default Setting

a

a. Unless preset by factory.

Min Output Voltage (V)

0.000 to

20.000

0.000

Max Output Voltage (V)

0.001 to +20.000

12.000

Min Output Current (A)

(see page 7)

0.000

Max Output Current (A)

(see page 7)

100.000

V-V Mode Input Range (V)

10.000 to +10.000

0.000 to +10.000

Stability Setting (%)

0.0 to 100.0

0.0

Coil Constant (kG/A)

0.001 to 999.99999

0.0

PSwitch Current (mA)

0.1 to 100.0

10.0

PSwitch Heated Time (sec)

5 to 120

15

Current Limit (A)

Min Output Current and

Max Output Current

80.000

Display Contrast (%)

0 to 100

80

Voltage Limit (V)

0.001 and

Max Output Voltage

2.000

Ramp Rate (A/sec)

(see page 7)

0.100

Programmed Current (A)

Current Limit

5.000