Ohaus CT Series ELECTRONIC BALANCE User Manual

Page 69

Advertising
background image

28

Stability Range

Stability Range

Stability Range

Stability Range

Stability Range

The stability range specifies how much a
displayed weight may change while the
stability indicator (unit indicator) remains
ON. When displayed weight changes
beyond the allowable range, the stability
indicator turns OFF indicating an unstable
condition. Portable

Advanced

balances

permit you to select one of four stability
ranges (in divisions) as shown in the table.

NOTE

NOTE

NOTE

NOTE

NOTE: When the RS232 interface is

configured to print stable data only, the

stability range also governs data output.

Displayed data will only be output if it is

within the selected stability range.

To view or change the stability range:

1.

Access the Stb submenu to display
the current setting.

2.

Press

to display the current set-

ting.

3.

Press

to change the setting.

4.

Press

to accept the displayed

setting, when released, Stb will be

displayed again.

STABILITY RANGE

STABILITY RANGE

STABILITY RANGE

STABILITY RANGE

STABILITY RANGE

.....5d

Smallest range: stability indi-
cator is ON only when dis-
played weight is within .5 divi-
sions.

1d

1d

1d

1d

1d

Normal range.

Normal range.

Normal range.

Normal range.

Normal range.

2d

Increased range.

5d

Largest range: stability indi-
cator is ON with displayed
weight changes up to 5 divi-
sions.