3 correcting mistakes – YSI BOD Analyst Pro User Manual

Page 64

Advertising
background image

YSI BOD Analyst Pro

®

User Manual

64

7.3 CORRECTING

MISTAKES

If you need to correct a Final DO reading that was stored you can retake Final DO readings just
as you were able to retake Initial DOs. The program allows you to correct the Final DO in the
Bench Sheet providing that the incubation period is between 4 and 7 days.

To do this after completion of Finals load the “read” batch into the Bench Sheet and select Read /
Retake Final
. Select to read All Bottles or Select This Bottle. If you select This Bottle then
click on the bottle of interest on the Sample Name row to take the reading. The program will
highlight the Sample Name to help verify that you selected the correct sample. The Windows
Take Reading display will reappear and if using the YSI 5000/5100 the instrument display will
change. ACCEPT/CONFIRM the new reading just as before to store the new DO value.

















As shown in the above figure you highlight the bottle you would like to read again (B#12) and
then click Read / Retake Final / This Bottle from the main menu. Read in the new value and
click Apply Changes at the bottom right of the Bench Sheet. A revised Bench Sheet appears.