YSI BOD Analyst Pro

®

User Manual

89

Stabilize timeout
This time in seconds is the max time the program will wait for the reading to become stable. If the
reading has not met stability criteria by this time, the program will store the reading anyway.
Legal values are 5 through 500.
The default value is 300.

Sample hold time
This is the hold time in seconds for Samples other than blanks.
The hold time in combination with Sample change is one criterion related to stability settings.
For a more detailed explanation see the “Auto-stabilization” section above.
Legal values are 1 through 300.
The default value is 10.

Sample change
This is the DO change in mg/l that is allowed during the hold time for a stable reading.
This field is for all Samples other than Blanks.
Legal values are 0.01 through 20.
The default value is 0.05.

Blank hold time
This is the hold time in seconds for blanks only.
For a more detailed explanation see the “Auto-stabilization” section above.
Legal values are 1 through 300.
The default value is 5.

Blank change
This is the DO change in mg/L that is allowed during the hold time for a stable reading.
This field is for Blanks only.
Legal values are 0.01 through 20.
The default value is 0.02.

Initial Low Temp Limit
If the temperature of your Initial DO is below this temperature (

C) the program will display a

warning message in the Take Readings window and (if using the YSI 5000/5100) on the
instrument display. You must ACCEPT/CONFIRM the condition before the reading is stored.
Legal values are 0 through 99.9.
The default value is 16.

Initial High Temp Limit
If the temperature of your Initial DO is above this temperature (

C) the program will display a

warning message in the Take Readings window and (if using the YSI 5000/5100) on the
instrument display. You must ACCEPT/CONFIRM the condition before the reading is stored.
Legal values are 0 through 99.9.
The default value is 24.