YSI IQ S ENSOR N ET DIQ/S 182 XT User Manual
Page 138
Technical data
System 182
9 - 2
ba76022e01
12/2011
Mechanical
construction
Ambient conditions
Electrical data
IQ S
ENSOR
N
ET
(line power version)
Enclosure material
Polycarbonate with 20 % glass fiber
Weight
Approx. 0.7 kg
Type of protection
– IP 66
– corresponds to NEMA 4X
The IQ S
ENSOR
N
ET
Universal
Transmitter is not suitable for Conduit
Connection
Temperature
Operation
- 20 °C ... + 55 °C (- 4 ... 131 °F)
Storage
- 25 °C ... + 65 °C (- 13 ... 149 °F)
Relative humidity
Yearly average
≤ 90 %
Dew formation
Possible
Site altitude
Max. 2000 m above sea level
Power supply
Nominal voltage:
100 ... 240 VAC ± 10 %
Frequency:
50/60 Hz
according to
DIN IEC 60038
Line power connection:2 pin, N and L
Line cross-section of mains connection:
Europe:
1.5 ... 4.0 mm
2
USA:
AWG 14 ... 12
Fuse rating on the operator side: max. 16 A
Protective class
II
Overvoltage
category
II
Power consumption
max. approx. 12 W