Error elimination

3.5.1 PROFIBUS error elimination

Data transmission

between the bus master

and IQ S

ENSOR

N

ET

is

faulty

The PLC does not

contain any plausible

input data

Here you will find causes and actions to take of errors con-
cerning the PROFIBUS and Modbus communication only.
General errors of the IQ S

ENSOR

N

ET

system are dealt with

in chapter 10 W

HAT

TO

DO

IF

....

Cause

Remedy

– Incorrect wiring

– Check/change the connections

(see section )

– Use checklist according to sec-

tion 3.3

– Incorrect setting of the

address

– Check/change the setting of

the address (see section 3.2.1)

– Incorrect protocol

– Check the version of the GSD

file

– Adapt the protocol

– IQ S

ENSOR

N

ET

defective

– Send IQ S

ENSOR

N

ET

to YSI

Cause

Remedy

– Input data and output data

are not consistent

– When programming the PLC,

define input data and output
data as consistent over the
entire data length

– The data interpretation of the

PLC is not correct

– Observe the data alignment of

the PLC data representation. If
necessary, exchange the
high-order and low-order bytes
word by word