8 technical data, 1 measurement characteristics, Technical data -1 – YSI NiCaVis 701 IQ NI User Manual

Page 49: 1 measurement characteristics -1, See chapter 8 t, Echnical, Data, Chapter 8 t, Ee chapter 8 t, 8technical data

Advertising
background image

NiCaVis 70x IQ NI

Technical data

8 - 1

ba76132e01

03/2013

8

Technical data

8.1

Measurement characteristics

Measuring principle

Spectrophotometric absorption measurement; integrated micropro-

cessor electronics, shielded 2-wire connection for

power and data transmission.

Wavelength range

200 - 390 nm

Measuring gap (optical

layer thickness)

Nitrite/nitrate

measurement

1 mm gap width

NiCaVis 701 IQ NI

1 mm

NiCaVis 705 IQ NI

5 mm

Measuring

location

Measuring mode

Measuring range

Resolution

Inlet

NO2

0.0 ... 120.0 mg/L

0.1 mg/L

NO2-N

0.00 ... 30.00 mg/l 0.01 mg/L

NO3

0.0 ... 300.0 mg/L

0.1 mg/L

NO3-N

0.00 ... 60.00 mg/L 0.01 mg/L

Activation

NO2

0.0 ... 120.0 mg/L

0.1 mg/L

NO2-N

0.00 ... 30.00 mg/l 0.01 mg/L

NO3

0.0 ... 300.0 mg/L

0.1 mg/L

NO3-N

0.00 ... 60.00 mg/L 0.01 mg/L

Outlet

NO2

0.0 ... 300.0 mg/L

0.1 mg/L

NO2-N

0.00 ... 75.00 mg/L 0.01 mg/L

NO3

0.0 ... 750.0 mg/L

0.1 mg/L

NO3-N

0.0 ... 150.0 mg/l

0.1 mg/L