Commissioning

NiCaVis 70x IQ NI

3 - 10

ba76132e01

03/2013

Measuring location

NiCaVis 701 IQ NI

:

Activation

Inlet

Outlet

NiCaVis 705 IQ NI

:

Outlet

(permanently set)

Measurement location or application of

the sensor.

The possible measurement loca-

tions are displayed depending on the

currently set measuring mode.

The sample matrix on which the cal-

culation of the measured value is

based changes with the location of

the sensor.

When the measurement location is

changed, all virtual sensors are

deleted.

Measuring range

-

Display of the measuring range

(see chapter 8 T

ECHNICAL

DATA

).

Cal - # raw value 1

Value pairs of the user calibration (see section 4.3.2).

Note: Leave those values unchanged during the initial commission-

ing.

Cal - ref. value 1
Cal - # raw value 2
Cal - ref. value 2
Number of N-sensors

Number of C-sensors

1 ... 2

0 ... 3

Apart from the main sensor, up to 4

more sensors can be displayed.

The settings for the additional sensors

are done in the menus for the respec-

tive sensors.

Zero adjustment

Factory

User

Setting of which zero adjustment is

used in the sensor.

Factory zero adjustment or

Zero adjustment carried out last

Signal smoothing

On

Off

Switching on/off the signal smoothing.

For more details, see

Response time

t90

.

Menu item

Settings

Explanations